Combined

http://www.youtube.com/watch?feature=player_embedded&v=3KpdLJiveE0

The combined test runs a creep test, then immediately transitions into an SRL test followed by dynamic loading. The sample thickness is updated between tests so that for example, the SRL test increments 10% of the creep test compressed thickness. On saving the data, Ey, and G* modului are calculated and placed into the CSV data file.

Parameters

See parameters for creep, SRL, and dynamic tests.

Known Issues

In transitioning between tests, erratic jumps occasionally occur, the cause of which has yet to be determined, but it may relate to what motor direction a given test initializes to.